C-Arm Compatible Surgical Imaging Table 1. General Requirements • Table shall be fully motorized with height, longitudinal, lateral, Trendelenburg, reverse Trendelenburg, and tilt adjustments. Must have float capabilities. • Minimum height shall be ≤28 inches from floor to tabletop. • Weight capacity shall be ≥500 lbs. • Tabletop shall be radiolucent with aluminum equivalence ≤0.4 mm. • Table shall be compatible with standard mobile C-arms (e.g., GE OEC series). • Battery-powered operation providing ≥8 hours of continuous use • Construction shall include stainless steel or carbon fiber components with smooth, nonporous surfaces suitable for surgical cleaning. • Table shall feature proportional hand and foot controls for motorized motion and floating tabletop functionality. 2. Dimensions and Mobility • Tabletop length ≥80 inches; width ≥24 inches • Longitudinal travel ≥35 inches • Lateral travel ≥10 inches • Trendelenburg and reverse Trendelenburg ≥±15° • Locking swivel casters with antistatic, floor-safe wheels 3. Accessories • Side rails compatible with surgical attachments • Radiolucent arm boards, straps, and footrest included 4. Safety and Compliance Requirements • CE Mark or FDA 510(k) clearance • Meets IEC 60601 electrical safety standards • Shall include emergency stop and battery indicator 5. Additional Requirements • Bid must confirm compliance with the listed specifications. Bidders are encouraged to include detailed technical sheets and demonstrate the equipment’s capability to meet veterinary-specific requirements, particularly for small animals. • Extended warranty to be negotiated with hospital after purchase. • On-site set-up by the manufacturer • On-site training by the manufacturer