Specifications include, but are not limited to: The required goods to be provided are Victor Niva Multilabel plate reader instrument along with a 4-year Service plan. Indicate Pricing Sheet, Exhibit C. Victor Nivo Multilabel Plate Reader Instrument # HH35000000 Base model F with the Time-Resolved Fluorescence (TRF) Fluorescence Polarization (FP) option # HH35000300 Victor NivoTM - Dichroic D400 unframed #HH35001971 Victor NivoTM Filter 615/8 Set # HH35000935 Victor Nivo TM Filter 320/75 Set # HH35000947 Requirements Filter storage capacity for up to 32 filters Ability to run Ricin assay. Reader certified to run Ricin assay. Delivery within 10 weeks of Purchase Order receipt. Installation of the Victor Nivo, Provide a virtual recorded training session, enabling end users to utilize the instrument for its intended purpose. Service Plan 4-years. Response Time - Malfunctioning equipment on-site response time must be within 3 business days. Emergency Breakdown visits – Unlimited, on-demand, on-site service visits for repairs as needed. Repair Service Parts – Included. All parts used for repair or replacement must be OEM (Original Equipment Manufacturer) parts. No refurbished parts may be used. Travel to our location – Included. Repairs must include diagnosis and correction of malfunctions and failures. Preventive Maintenance Visits – 1 or 2 Annually Technical and Remote Support – Regular hours Operational Qualification Service Discounts – 10% Hardware and Software Upgrade Discounts – 15% Installation Service Discount on Training – 15% Due to the critical public health nature of this equipment, no aspect of this statement of work may be subcontracted. Service engineers must be qualified, and factory trained.