Specifications include, but are not limited to: The testing kits shall contain materials to perform at minimum forty-eight (48) tests. The kit shall consist of sensors, tubes of treatment reagent, capillary tubes and plungers, transfer droppers, two (2) levels of quality control, and a calibration button. • Power requirements – 1.5 volt alkaline or lithium batteries (Four (4) each) or 100-240 volts AC, 50- 60 Hertz (Hz), 3.3 volts Direct Current (DC), 1.2A • Sample volume – 50 Microliter (μL) • Sample Type – capillary whole blood (Ethylenediaminetetraacetic Acid (EDTA) or herparin anticoagulants) • Test time – Three (3) minutes • Test method – electrochemical with disposable sensors • Quality control – Two (2) levers, room temperature stable ( including test kit) • Reportable range – 3.3 – 65 Micrograms per Deciliter (μg/dL) • Calibration – electronic calibration button