Specifications include, but are not limited to: Qty: 3 - Glove Box Systems Specific requirements for each glove box are: Glovebox 1 • Dual Workstation (four gloves operated side-by-side) • 2500 mm length assembled from two 1250 mm modules (elevator size restriction) • Integrated Gas Purification system (one purifier) • 90 cubic meter/hour ventilator • Two 0.3 micron dust filters (one for inlet, one for outlet) • PLC Controller and HMI • Automated Regeneration • Large and Small Antechambers • Automated Purging (Large Antechamber) • Expandable Sidewall • Replaceable Sight Glass • Target Purity Level: <1ppm Moisture , <1ppm Oxygen • Housing & Antechamber must be made of 304 Stainless Steel • 2 Sight Glass with 2 Glove Ports (Each) • 4 Butyl Gloves