Specifications include, but are not limited to: Bid Specifications for the AS-AP autosampler Autosampler must be tested and certified to be compatible with and capable of being integrated with the existing Dionex ICS3000 Ion Chromatograph. Must be controlled by existing Chromeleon software version 7.1 or latest version. Must include 3 vial trays that have range capability of 0.3-1.5mL, 40 positions. Must include 100 pack 1.5mL glass vials with caps and septas. Entire sample and fluid path, including rinse lines, must have inert, nonmetallic PEEK components to ensure solvent compatibility and metal contamination-free Chromatography. Must have the following injection volume range: 1-100uL in 0.1uL increments. Must have sample overlap capabilities to start sample preparation while the instrument is acquiring data with real time access. Must inject in less than 30 seconds for a 5uL full loop. Must have support for the following types of injection modes: Full loop, Partial loop, Limited Sample, and Capillary