Specifications include, but are not limited to: X-ray system must be digital and must support both wired and wireless operation System must include high-definition DR panel: Approximate sizes 10” x 12” System must be compatible with the Golden X-ray generators System must be self-contained for power, capable of at least 3-hour battery powered operation Must have True Long-Exposure capability, able to produce a one-pulse X-ray from 30ft. System must include Auto-Stitch image stitching capability Must have software controlled generator firing and sequencing Includes rugged carrying case System must meet the Ingress Protection rating of IP65 (Panel at IP67) for dust, dirt, splash; and a 3-foot drop shock tolerance System must have short range and long range modes of operation while operating wirelessly DR Imager Panel must have Titanium Backscatter Protection System must support date and time stamps on each image shot Cold boot to first image is approximately 60-75 seconds with second exposure full cycle at approximately 8-12 seconds Software must have 3D capability System must have been evaluated and approved for all explosive ordnance disposal use by the Military Technical Acceptance Board, Naval Surface Weapons Center