Work will include (but not limited to) hood and transport velocity, general conditions of the system (hood, ducts, fan system, and associated equipment), identifying recommended repairs, and provide cost estimates. Fume hoods are to be certified for testing and repairs according to all required State and Federal regulations and requirements. All hoods tested are to be documented, and a survey summary is to be provided to the Office of Facilities and Operations. The summary should include the following information (in an Excel spreadsheet) and provided annually: • Campus • Date • Building • Room • Sash Parameters • Measured Face Velocities in FPM (feet per minute) at the following locations: o Upper Left o Upper Middle o Upper Right o Lower Left o Lower Middle o Lower Right • Average Face Velocity • Air Flow rates in ml/sec (milliliters per second) and CFPM (cubic feet per minute) • Class • Hood Type (VAV/CAV)...