2.1 MANUFACTURERS A. Subject to compliance with requirements, products that may be incorporated into the Work shall be as follows: 1. Basis of Design (BOD): 70cm-78cm bore, 16cm AFOV CT with high sensitivity, 256 slice CT, fixed installation, required cable connections, chiller and cooling hose lines. Manufacturer: Siemens. a. Scalable, fixed installation, high-sensitivity, CT system, 256 slice. b. Imaging tools, including image analysis, optimization, motion correction and gating. c. Auto positioning tools. d. Imaging optimization to reduce dose, improve resolution and reduce noise and other artifacts. e. Camera system to allow CT scanner from the control room. f. Necessary tools provided for method of install and full installation of equipment. 2.2 STANDARD SYSTEM HARDWARE A. Two MX SIGMA X-Ray tubes and generators. B. Standard dual head technology. C. One dose shield. D. Two detectors (Min: 30 mm x 30mm) E. Two 100 KW generators. F. Cooling water system or connection to public water system. G. CT table, standard size, 400 lb.load. H. Up to 140 FPS image reconstruction. J. Dual 19” FSM with dual display. 2.3 CT PERFORMANCE SPECIFICATIONS A. X-Ray Tubes: Liquid bearing tube enables 0.28 second gantry rotation speed for routine use. B. Thermal ratings: 1. Maximum Anode Heat Content (Reference: IEC 60613) 2. Maximum X-ray Tube Assembly heat content: 7.7MJ (10.8 MHU) 3. The maximum anode heat capacity: 5 .0 MJ (7 .0 MHU) 4. Anode heat dissipation: 1070 KHU/min (13.2 kW) C. Distance of focal spot to detector: 95 cm D. Maximum power: 200kW E. mA: mA range at 120 kV: 10 to 600 mA, 5 mA increments