Beckman Coulter Allegra X-30 IVD Benchtop Centrifuge, Item or Catalog Number B05796, or equal * Note: Brand/Make and catalog numbers given for reference only and are NOT to be construed as the required brand/make and catalog numbers. Centrifuge shall be installed in the Sheriff’s Department’s Toxicology laboratory. The centrifuge shall be used to separate and isolate substances in blood and urine samples associated with DUID (Driving Under the Influence of Drugs) and sexual assault cases. CERTIFICATION REQUIREMENTS (NO EXCEPTIONS): Direct drive brushless motor or equivalent Centrifuge dimensions: approximately 18.1(W) x 21.7 (D) x 14.0 (H) closed, 31.0 (H) opened (inches) Weight: approximately 100-110 lbs. Power requirements: 120V, 6A, 60Hz The control panel shall be on the front of the instrument and include system keys, programming keys, and digital displays. The control panel must have an imbalance light that flashes if the centrifuge is out of balance. Must have RPM/RCF/Rotor, Time, and Acceleration/ Deceleration digital displays. Ambient temperature range: approximately 4 to 35 degrees Celsius Humidity restrictions: approximately <80% (noncondensing) Noise level (3 ft.) in front of instrument: approximately 65-70 decibels Maximum heat dissipation into room under steady-state conditions: approximately 1600-1675 Btu/h Speed: Set speed: up to approximately 16,000 rpm (in 100- rpm increments Speed control: actual rotor speed, approximately +/- 50 rpm of set speed