Description: Level 3, ISO Accredited, On-Site Service, ISO 17025 (4) As Found, (10) As Left: High, Middle, Low Approximately 177 Pipettes calibrated twice annually onsite. *each equals number of Pipettes Range is .2ul - 1000ul and 8 and 12 channels. -Vendor Should be ISO/IEC 17025:2017 Accredited -Reports on pipettes that failed calibration need to include "As Found" data as well as "As Left" data The Preventive Maintenance includes: -Cleaning the piston -If needed; re-greasing, replace seal, replace o-ring -Passed pipettes must be returned with individual calibration certificates reflecting the testing results and an indication of pass or fail. Calibration stickers must be adhered to each pipette on paper and/or digital format indication calibration date on an annual calibration