Specification Requirements: Frame: All physical frame dimensions, if listed in Vendor documentation in metric units, can be converted and rounded to whole English units to demonstrate compliance with these specifications. The maximum test speed can be converted to the nearest whole English units. The minimum test speed can be converted to the nearest five decimal English units. System power/operation: Electromechanical Meet Buyer’s Maximum power voltage: 120 Volts AC at 60 Hz, Single Phase Loading controlled by dictated load, displacement, or strain Ability to test specimens up to 12 square inches Load Capacity: 20,000 lbf Maximum test speed: 0.2 inches per minute Minimum test speed: 0.002 inches per minute. Load measurement accuracy: ±0.25%; Included Hardware and Devices Shear box assembly Filler block 12 inch x 12 inch porous stones Data cable Compaction table 12 inch square load pad...