Specifications include, but are not limited to: X-Ray Panel Wireless, glass free, x-ray panel. Image size no less than 21 inches measured diagonally with approximately 85% of the panel actively used for detection. DR panel drop tested for durability of at least three feet (36 inches) DR panel has a multi directional connection. DR panel able to operate in both a wireless and wired configuration, with cabling connectors meeting or exceeding MIL-DTL-38999 standards. Two rechargeable batteries with charging platform. Communication/Control System will include two tablet style touchscreen computers, to control and analyze the produced x-ray image. Each tablet computer will have a resolution of at least 1900 X 2100. Each tablet computer will be ruggedized (MIL-STD-810G) Each Tablet will run a Windows operating system. Each tablet computer will be approximately 10 inches in size (measured diagonally). Each tablet computer will have a screen brightness to be operated in direct sunlight, meeting or exceeding NIT 1500 standards. Each tablet will have HDMI output. System will be operated by a secure wireless mesh communication system, with two transceivers, that can attach to the controller tablet computers Each tablet will be able to control and display x-ray images. Another transceiver will be able to attach to a Golden Engineering X-ray generator and be able to be powered by the X-ray generator’s 20 volt Dewalt battery Transceivers will have a line-of-sight range of approximately 0.4 miles.