Specifications include, but are not limited to: Equipment Requirements - Biochemical Analyzers: By no later than ten (10) calendar days after the state agency authorizes the contractor to proceed with services, the contractor shall provide and install two (2) biochemical analyzers in the laboratory in accordance with the following: The contractor’s biochemical analyzers must be new within one year of manufacture date. All biochemical analyzers provided by the contractor must have the ability to run both oral and urine samples, and must have a service agreement to cover all repairs, parts, and maintenance at no additional charge to the state agency. The biochemical analyzers shall meet or exceed the standards listed below. In the event the state agency’s standards change, the Division of Purchasing will amend the contract to revise the following standards. Random or stat access, specimens or test selective analysis with bar code reader capability. Analysis rate of a minimum of one thousand six hundred (1,600) photometric tests per hour by each biochemical analyzer. Both biochemical analyzers must be able to fit into a nineteen foot (19’) by six foot (6’) space. Rack handlers must be automated...