Specifications include, but are not limited to: A. The Contractor shall provide intact drug testing kits which incorporate collection and testing for the detection of drugs of abuse in human urine and oral fluids. The kits will provide a reader, which will be included at every site at no additional cost (about 15 readers total-statewide). B. The Testing Kits shall test and meet the following criteria: 1. Kits shall be fully integrated, round (minimum height 3.15 inches), upright (no flat side), one step device for commonly abused drugs in urine with an inner lock for transport (to prevent leakage). Devices cannot require tipping, tilting, card insertion, or key turning. 2. The integrated test strips must provide all of the following drug testing assays: • Amphetamine 500 • Benzodiazepines 300 • Cocaine 150 • EtG (alcohol) 500 • Fentanyl 500 • Methadone (MTD) 300 • Methamphetamine 500 • Opiates 300 • Oxycodone 100 • THC/Cannabinoids (THC) 50 For the oral fluid testing, it must be self-contained and not require the use of a pipette. Shipping will be provided by USPS priority or UPS only and must be able to test for Amphetamine, Cocaine, Methamphetamine, Opiates, Phencyclidine, and Alcohol Contractor shall list all other drugs which can be included into the drug test kit format. (In order to provide deterrence, donors must be kept guessing. Contractor should be able and willing to provide different panels in order to achieve this goal.) The Contractor should be able to provide testing for alcohol, spice, ketamine, and fentanyl. 3. Contractor will provide pricing both for kits with and without the capability to do specimen validity testing (SVT). SVT shall minimally test for pH, Oxidants and Specific Gravity or Creatinine and shall be an integrated part of the test kit. 4. Each device shall have an affixed temperature strip. 5. The test cup reader must be able to read the urine device within 60 seconds. Must be capable of reading both the 6 and 10-panel test cups. The device must have a digital touchscreen and be easy to operate. Must export both .csv and .pdf files. It must provide digitally archive-able data. The readers must be delivered prior to the first order. 6. Contractor shall describe in detail their customer support component including the qualifications of assigned personnel. 7. Each device shall be in a sealed foil pack that includes a desiccant pack, a lot number, an expiration date, and the specific tests that are part of the panel. 8. All kits shall have, upon delivery, a minimal shelf life of one year. 9. Each drug shall have its own control line and one drug per strip is preferable. Preference shall also be given to products with a tear/pull-off results window to provide additional security for testers. 10. The six or ten drugs will constitute the main panel and should have a combined accuracy rating of at least 96% when compared to GC/MS.