Specifications include, but are not limited to: Clemson University seeks bids for a Liquid Chromatograph-Triple Quadrupole Mass Spectrometer. The Department of Environmental Engineering and Earth Sciences (EEES) at Clemson University seeks to procure a Liquid Chromatograph-Triple Quadrupole Mass Spectrometer. Initial research projects and interests reside in the exploration and characterization of polyfluoroalkyl substances (PFAS), algal toxins, and cannabinoids. Minimum Requirements: Compatible with data acquisition and processing software that must allow for the seamless method transfer between our existing Agilent LCMSMS system and the new LCMSMS instrument for the purpose of research continuity of ongoing projects. Sensitivity in positive ion mode such that a 1 pg injection of reserpine injected onto the analytical column can be quantified at the transition m/z 609.3 to 195.1 with S/N > 850,000:1. Sensitivity in negative ion mode such that a 1 pg injection of chloramphenicol onto the analytical column will be quantified at the m/z 321.0 to 152.0 transistion with a S/N ratio > 850,000:1. Instrument detection limit (IDL) in positive ion mode of less than 3.5 fg for a 10 fg injection of reserpine at the m/z 609.3 to 195.1 transistion. Instrument detection limit (IDL) in negative ion mode of less than 4 fg for a 10 fg injection of chloramphenicol for the m/z 321.0 to 152.0 transition.