The Bid must meet or exceed the following Specifications. ▪ Include a recirculating water chiller if required for ICP-OES System and chiller must utilize existing 120V outlet. ▪ Must be capable of simultaneously analyzing axially and radially a wavelength range of at least 167-785nm. ▪ Must include an autosampler designed to meet the requirements of a medium to highthroughput lab. ▪ Meets or exceeds requirements for analysis of SW-846 EPA, EPA 6010B, 6010C, and EPA 200.7. ▪ Hardware: Computer with Windows 10 Operating System (COT does not support Windows 11), 24-inch monitor, color laser printer, software, and adequate memory to run instrument and process data. ▪ Software: Must be capable of exporting data to excel or .csv format. ▪ System must have dual view capability for analysis of easily ionized elements. ▪ System must have single step torch alignment capability. ▪ Installation kit must include the following: wavelength checkout solutions, tubing (pump, waste, sample introduction, gas line), waste container, and consumables to install, start up and analyze upon installation. ▪ Must fit existing footprint of 144" wide x 27" deep. ▪ Installation must be completed by the Seller. ▪ 3-4 days in-house training and method development QC checks must be included. ▪ System must be capable of meeting a Cd LOQ <1.0 ppb, Ag LOQ <0.5 ppb, and a Cu LOQ <0.5 ppb. ▪ Guaranteed 72-hour on-site technical support and unlimited telephone technical support. ▪ 12-month comprehensive warranty. ▪ System must utilize existing exhaust system and must not be required to exceed 200 cfm.