Specifications include, but are not limited to: Baker BSC serial number 120081 in room 2313. Disconnect Canopy/Thimble connection from BSC. Disconnect thimble connection from exhaust duct work. Install exhaust grill where thimble connection was removed from the buildings exhaust ductwork. Replace glass BSC sash with new Baker BSC sash and bring BSC back to normal working order. Certify BSC. Baker BSC serial number 112108. Gas decontaminate BSC. Purchase, ship, and replace the Supply and Exhaust HEPA filters. Certify BSC. Nuaire BSC serial number 57309 ABN Gas decontaminate BSC. Disconnect BSC from Ductwork Dispose of BSC. Install exhaust grill where previously connected to the ductwork. Nuaire BSC serial number 57307 ABN Gas decontaminate BSC. Disconnect BSC from Ductwork Dispose of BSC. Install exhaust grill where previously connected to the ductwork. Move Labconco BSC located in Lobby to nearby room.