Specifications include, but are not limited to: • General: High-performance sterilizer (autoclave) specifically designed for sterilizing or decontaminating laboratory items of many sizes and types, with programmable treatment options. • Basis-of-Design Product: Primus Sterilizer Company, Model PSS5-B. 1. UL listed and/or approved electrical devices and components, bearing NEMA MG 1, IEC or other recognized International ratings, as appropriate for the use intended. Low-voltage DC outputs external to the electrical box. • Cabinet Enclosure: Yes. • Mounting: On floor, free-standing. • Chamber Interior Dimensions: 26 inches wide by 26 inches high by 39 inches front to back (660 mm wide by 660 mm high by 991 mm front to back). • Bio-seal: Yes. • Configuration: Configuration: Double Door (pass through), recessed into one wall. • Doors: Number, type, and construction indicated. 1. Number of Doors: Two. 2. Hinged, manual-type. 3. Door handing. 4. 329 TB Waste Left hinged, 2 doors 331B BSL-3 Waste Left hinged, 2 doors 5. Gaskets: Manufacturer's standard for type door specified, manufactured to withstand temperatures and pressures generated. • Operator Controls: Digital type, with touch screen controls and multi-level menus. • Control and Monitoring System: Controller, gauges, and wiring. 1. Controller: Microcomputer based controller monitoring and controlling all aspects of equipment cycle and process operations. Provide battery backup of microprocessor memory. Controller to ensure fully automatic operation through all equipment cycles.