Specifications include, but are not limited to: Mississippi State University College of Veterinary Medicine requests bids for a new Overhead Tube Crane for Small Animal X-Ray applications. This unit must be fastened and suspended from the ceiling and include the components that make up the x-ray system such as the generator, tube, exam table, detector, and computer with software to operate the equipment. Overhead Tube Crane Specs: ● One pair of longitudinal rails 14 feet in length to cover the total travel of the room. ● Tube Rotation capable of performing cross table projections. ● Vertical telescope travel range of at least 70”. ● Longitudinal travel range of at least 130” ● 65kw high frequency 3 phase generator. ● Collimator with manual controls, bucky light, and measuring X-Ray Tube ● Maximum Tension – 150 Kvp ● 90 degree cable arms 65kW Generator Specifications: • 150Kvp • Digital interface for integration with digital radiography systems • Anatomical programming with at least 700 programmable technique selections. • Tube protection circuitry • Dual Speed starter • Self-diagnostic circuitry with error codes for troubleshooting. • Deluxe heavy-duty manual cassette tray. • Differential SID Sensor • Table top hand control that controls all table functions including elevation and the 4-way table top. • 17”x17” grid