Specifications include, but are not limited to: Purchase Description The radiation measurement meter must able to: 1. Take measurements in regular radiographic, fluoroscopic, and CT machine settings. 2. Measure kV and kVp from 40 to at least 150 3. Display kV waveform 4. Measure exposure time from 1 ms to 999 s 5. Measure exposure pulses 6. Measure dose and dose rate in both standard and SI units 7. Display dose rate waveform 8. Measure half-value layer and total filtration 9. Measure light intensity in area as small as 10 mm 10. Display photopic response curve 11. Measure CT dose with a sensor having an active length of 100 mm. Sensor must be able to fit within Ludlum CT phantoms model L-441, Model L-451, Model L-491 12. Sensor capable of taking leakage and scatter measurements for both Air Kerma (in standard and SI units) and Dose Equivalent. 13. Meter and sensors should have automatic environmental compensation 14. Must be Bluetooth capable 15. Must be able to export data to Microsoft Excel