Specifications include, but are not limited to:Device must: Hand-held Use Raman Spectroscopy Data output formats: CSV, Excel, SPC, PDF An operating temperature of -14F to 122F (-10C to + 50C) Have heroin, K2/Spice plant material, and Fentanyl (all 3 forms) testing capability Have mixed chemicals analysis capability A library consisting of 213 or more controlled substances and 135 or more cutting agents or precursor chemicals Clearly identify and display the identity of the chemical(s) identified A single test for multiple controlled substances that provides clear, definitive results for presumptive identification with no user interpretation Scan directly through plastic or glass to minimize contamination, reduce exposure and preserve evidence Capture all scan results, including time-and-date stamp and system self-check results to help support prosecution Compatible with Companion PC TruNarc admin software Not require any scheduled maintenance or calibration Self-check function that reports a Pass or Fail result to insure proper operation Scan data saves automatically with simple report generation upon USB download Closed library, ensuring the integrity of analysis results Include lifetime controlled substance library updates Reachback support service, providing secondary analysis of ‘Inconclusive’ Scan results Include integrated online training and available onsite training Rechargeable battery with a minimum of 10 hours of operational time Capable of being charged through a standard wall power outlet, car charging port or PC USB connection Boot up from cold start in 20 seconds or less and require no calibration or warm up time