2. SPECIFICATIONS
. Minimum of three low temperature sterilizers shall have a minimum individual dimensions of
Cabinet (DD): 33 x 40-15/16 x 72” (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38-
5/16 x 72” (838 x 973 x 1829 mm).
o Mounting single door, cabinet
o System installation requires no plumbing, ventilation or air
o supply – only a dedicated electrical connection is needed. A power cord is supplied for
this connection
o The facility space in which the vehicle can occupy is approx. 35’ x 110’.
. Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive
materials because temperatures within the load currently do not exceed 55ºC (131ºF)
and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can
also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized
hydrogen peroxide sterilizers, either with or without plasma, are currently available in the
United States.
. Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area
with a ventilation system capable of at least 10 air exchanges per hour.
. VPRO MaX2 must support utility requirements:
o Current 16 AMP
o Minimum Circuit Breaker 20 AMPS
o Use NEMA L15-20R Plug
o Provide grouped or ganged circuit protection and disconnect for phase 3 power as
required by NEC and local codes.
o Individual power shutoffs near each machine for servicing.