The Oklahoma City Veterans Affairs Health Care System is seeking sources for a medical electrometer to perform tests performed by the Radiation Oncology department. The device is anticipated to have the following requirements: Minimum requirements Remote Access Full remote access through a web-browser Flash compatible Supports your TG-51 QA protocols Connectivity via WiFi, Ethernet or USB 3.0 Host Correction Factors Full remote access through a web-browser Flash compatible Supports your TG-51 QA protocols Connectivity via WiFi, Ethernet or USB 3.0 Host Library Functions Extended detector library Check source library Machine library API Functionality Large touch screen Desired specifications Connectivity Ethernet LAN 1 Gbit/s, Wi-Fi (802.11ac/b/g/n), USB 3.0 Host Dimensions (at or near) 249.7mm × 201.9mm × 123.4mm Current Range 400 fA 24 A Charge Range From 24 pC to 65 nC per pulse at a repetition frequency of 400 Hz Configurations TNC Triax, BNC Triax (Adapter) Zeroing 60 s (default) Bias Voltage Â} 50 V Â} 500 V, Â} 2 % Zero Drift Â} 0.25 % ( Â} 0.5 %, IEC 60731) Non-linearity Â} 0.25 % ( Â} 0.5 %, IEC 60731) Response Time < 1.5 s Long Term Stability Â} 0.2 % ( Â} 0.5 % over one year, IEC 60731) Repeatability < Â} 0.25 % ( Â} 0.25 % IEC 60731) Measurement Modes Manual, Automatic (beam triggered), Timed Resolution 0.1 fC Additional information This is a request for information and does not represent an order or any obligation on behalf of the VA nor does it commit the VA to a future request for quote (RFQ). Please provide the country of manufacturing for devices you would consider quoting if an RFQ is issued. If you consider your company a potential source for this type of device please provide the following information: Company name, Name of a point of contact including email address, Size demographics of company Questions? Deadline for questions 8/26 Point of contact: Christopher.delauro@va.gov