A24-007 Water Cooled Xray System 76 AMXG Request for the purchase of 4 ea.
Water Cooled Xray System to include the following:
Input Line Requirements:
Automatically adapts to input line voltage100-130Vac, 50/60Hz, 20 Amperes maximum.
200-250Vac, 50/60Hz, 10 Amperes maximum.
10 to 160kV, 0.5mA to 5.0mA (800 watts max.)
Constant potential, end-grounded anode, liquid cooled version.
X-Ray Tube Window: Beryllium 0.8mm (directional) 40° directional.
Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm).
aluminum at 160kV, 5mA.
Radiation Leakage: Less than 2.0 Roentgens per hour at 1 meter from the X-Ray tube target.
Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN125430.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336.
Ambient Temp: 100% duty cycle @ 120°F (49°C).
Duty Cycle: 100%.
Storage Temp: -30°F to 160°F (-35°C to 71°C).
Anode Cooling: Liquid coolant solution closed loop between X-Ray tube anode and cooling unit.
Liquid Cooling Unit Weight: 54lbs. (15.4kg) approx.
Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length.
Liquid Cooled: 29lbs. (13.15kg) X-Ray Control Unit: X-Ray Control Unit Weight: 26lbs. (11.8kg) approx.
Safety Devices:
Tubehead Pressure Relief Valve
Tubehead Thermal Cut-Out
Tubehead Pressure Gauge
Tubehead Low Pressure Cut-Out @ 25psi (1.72 Bar)
Coolant Flow Sensor (liquid cooled only)
Control Unit Safety Keyswitch
Microcomputer-based Self Diagnostics
Continuous Exposure Parameter Display
Standard Accessories:
Operation manual
Tubehead carrying case
Tubehead Cable- 100’ with strain relief
Extra key (1) for Control Unit Safety Lock
Power cable - 25’ with strain relief
Coolant hose - twin, 50’ with self-sealing terminations
(liquid cooled only)
Cooler power cable
GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user-friendly windows-based menu. Additionally, the GUI can be used as a diagnostic tool when the system is controlled via the front panel.