Question and Answer posted in attachments/Links.
MOD 0001
RFQ CLOSES ON 8/11/22 @ 1159PM
The Minneapolis VA Healthcare System Blood Bank Laboratory, 1 Veterans Drive, Minneapolis Minnesota, 55417, here in after referred to as the MVAHCS, has a need for seven Blood Bank centrifuges, which will be used to support the unique needs of Blood Bank. Various testing such as cell washing, ABO/Rh typing, antibody screens, antibody identifications, DATs and cross matching is performed with these centrifuges.
-
Required capabilities:
-
Maximum capacity of centrifuging 12 tubes, simultaneously
-
Buckets must hold 10x75mm tubes
-
Maximum RPM of 6,000
-
Maximum RCF of 4,146
-
Power requirements of 115 V
-
Dimensions not to exceed 9.5” H x 13” W x 15.75” D
-
Programmable presets available
-
60° swing-out rotor
-
Warranty:
-
A five-year manufacturer’s warranty at no extra charge to MVAHCS is to be included in purchase.
-
Delivery and Installation:
-
All new equipment will be shipped within 30 days of receipt of award. The contractor will make arrangements with MVAHCS for delivery and installations of equipment. The MVAHCS requests delivery terms of: FOB DESTINATION for all shipments, deliveries, installations, and training.
-
Deliveries shall be made to:
VA Medical Center, Bldg. 70
Warehouse (90D), Ref. PO#:____________
One Veterans Drive
Minneapolis, MN 55417
-
Acceptable delivery times are between 8 am and 3 pm CST, Monday through Friday, except Federal Holidays.
-
Installation will be completed by our Biomed department.