Specifications include, but are not limited to: Kansas State University, a Kansas state agency and public state university (“University”), for and on behalf of 05040 Division of Financial Services (“Department”), is accepting Quotations for one (1) Flow Cytometer with the following requirements: A bench-top high-performance 3 laser (violet, red and blue) Flow cytometer with 30-40 fluorescence detectors. The system must be easy to learn with a user-friendly software. Must have a minimum of 35 detectors. Must use high sensitivity coarse wavelength division multiplexing (CWDM) 16-channel semiconductor detector array for 405nm laser, 14 channel semiconductor detector array for 488nm laser and 8-channel semiconductor detector array for 640nm laser. Must use full spectrum unmixing/compensation. Must be a small footprint system (bench-top) model. Must offer a wide array of new fluorochrome combinations without reconfiguring system for each application (no filters needed). Must use vacuum fluidics. Must use spatially separated laser beams. Must use violet laser SSC. Must have an automated sample loader. Must use 22-bit 6.5 log decades. Must include instrument training by technical representative. Preferentially include up to 4yrs of extended warranty/service contract.