Specifications include, but are not limited to: Accuracy of the automatic length measurement: 0.03 mm Resolution over the entire measuring range: 2 nm Measuring range: ± 5000 μm Sample length: 0 mm to 52 mm Sample diameter: 12 mm (optionally 19 mm) Force range: 10 mN to 3 N in 0.2-mN steps Temperature accuracy: 1 K Temperature precision: 0.1 K Temperature resolution: 0.001 K Temperature stability (isothermal): ± 0.02 K Gas atmospheres: Inert, oxidizing under static and dynamic conditions Contains High-temperature tube furnace with exchangeable SiC heating element, prepared for exchangeable protective tube made of alumina or fused silica (for gas transit). Room temperature to 1650 degree C Type S thermocouple Should include the comprehensive software Should include Protective tube of Al2O3, for gas flow, complete, with support flange, maximum temperature 1650°C Should have Tube design sample holder kit of Al2O3, for sample lengths 0 ... 52 mm and sample diameters up to 12 mm, with holding tube, pushrod and sample thermocouple type S, including a set of sample supports for samples Ø 4 mm, 6 mm and 8 mm, - for 12 mm sample diameter no separate sample support is needed, maximum temperature 1600°C Should include the standard materials Should include Sample container from Al2O3 for dilatometer measurements on pasty and powdery substances, consisting Should include the on-site installation and training.