Specifications include, but are not limited to: Magnet Specs: • Filled magnet weight of 5,000 kg or less • Patient aperture at isocenter is greater than 69 cm • Homogeneity less than 2.0 ppm at a scan field of 50x50x45 cm • Maximum scan FOV greater than or equal to 55x55x50 cm • High order shimming capabilities: first (x, y, z) and second order (x2-y2, z2, xy, xz, yz) for improved patient specific shimming. • Integrated posterior surface coils in table with coverage of greater than 180 cm in length without repositioning patient • MRI table must be dockable, or have an integrated elevating trolley • Table/trolley patient weight limit > 240 kg • Weight of table/trolley for patient transport is less than 135 kg Imaging System Specs: • Gradient strength > 35 mT/m • In-plane slew rate > 155 mT/m/ms • Analog-to-digital image signal conversion in the surface coil, or on the table • Channel-less receive architecture (unlimited channels) • Minimum 2D TE < .28 ms (128 matrix) • Minimum 3D TE < .23 ms (128 matrix) • Minimum 2D TR < .80 ms (128 matrix) • Minimum 3D TR < .80 ms (128 matrix) • Patient-adaptive B1 shimming with 2 or more independent RF amplifiers • Wireless physiology triggering • Average power consumption < 20 kW/hr Surface Coils Specs: • Transmission of signal for Head/Neck/Spine & Body Coils by fiber-optics (not pins) • Head Coil with > 15 channels • NV Coil with > 20 channels • Body/Pelvis Array coil with > 32 channels • Small Extremity Coil with > 16 channels Surface Coils Specs (Continued): • pediatric coil 8ch with Internal dimension 23cm x 17.5cm for imaging small animals • Small (10 cm) Flex Coil with > 4 channels • Medium (15 cm) Flex Coil > 6 channels • Large (20 cm) Flex Coil with > 8 channels