Specifications include, but are not limited to: Chamber Volume 19.5 x 29.5 in x 29.5 in Capacity (minimum) 160 L / 5.65 ft3 Shelves 2 w/rack Loading Cart/Transfer Carriage None Door Type Hinged Single Door, Fully Jacketed Steam Generator Electric, 3-Phase, 480V, Built-In Mounting Cabinet, side enclosure panels / free standing installation Air Compressor Air Removal Pre-vacuum with standard water ejector (most applications) Programmable Pre-set and customizable Standard Programs Sterilization of liquids, gravity cycles, pre-vacuum cycles, air removal and vacuum leak (test programs) Cycles Gravity, Liquid, Pre-Vac Test Cycles Vacuum Leak Test, Bowie-Dick Tests Temperature range 105 to 137 (°C) / 221 to 278 (°F) Real-Time Cycle Data Including pressure, temperature, change of phase, alarms and a summary at the end of cycle Printer Alpha-Numeric, Thermal, Built-In Controls/Display Built-In, multi-function touch screen panel, built-in memory, internal battery backup, multiple access levels for added security Piping Stainless Steel Threaded Valves Air-Operated