Specifications include, but are not limited to: Perform annual inspection, testing and certification to include: a. Chemical Fume Hoods: i. In accordance with Scientific Equipment Furniture association (SEFA) standards, perform airflow testing and visual smoke testing. It is the vendor’s responsibility to be familiar with national standards that must be adhered to in order to “certify” that units are operating properly. b. One 4-foot wide radioisotope fume hood, model #5-0074-2L, hood #H-38. Testing and certification for this hood shall be completed both with and without flow through the HEPA filter (via separate exhaust ducts) 2. Replace as required, pre-filters, HEPA filters or exhaust filters on biological safety cabinets, BacTec instrument, and ductless fume hoods. 3. Provide decontamination services as required Biological Safety Cabinets and decontamination services of Biological Safety Level III Laboratory areas (Room Decontamination). Please refer to the NSF/ANSI Standard 49 for the Acceptable agent for the decontamination of the BSL3 laboratory. 4. Vendor shall be responsible for proper packaging of contaminated materials (IE. Filters) for disposal, to be handled by State of NH. 5. Replace Ultraviolet (UV) light as required. 6. Inspect and replace as required, HVAC in-line Bag In – Bag out Filters. These filters exhaust Biosafety Level III laboratories. There are four (4) units that are all the same model (BG1-1H1W-2GGF-304-D3) F a. Pre-filter M/N 24X24X2 (normal) b. Filter/Absorber M/N 24X24X11 ½ Actual c. Change out bag size – Pre-70CX72 HEPA – 84CX96 7. Set up separate accounts for each agency to include the proper agency name, service address, contact person and billing address.