Specifications include, but are not limited to: Mixed C/H/O/S/N Gas Chromatograph for CO2 Utilization Gas Characterization. Specifications for Mixed C/H/O/S/N Gas Chromatograph for CO2 Utilization Gas Characterization UND is undergoing a project with another University that is aimed at developing a novel method of conversation of fossil power plant flue gas CO2 to products. A crucial component of this project will be the accurate measurement of extremely varied gas streams continuously, with a large amount of required gas components required for simultaneous measurement. The instrument must meet the following specifications: • Automatic sampler of inlet gas feed stream into the unit is required, with minimal user input. This does not include manual injection via syringe, but a port for this ability is requested (bids will not rejected if no port is included). o Sampler must be able to automate in some capacity without a user present at times and able to take at least 10 samples at different times without user input. • Must be able to accurately measure (to below 1 ppm accuracy from up to 100+ ppm concentration) of each of the following gasses, specifically, as well as the following general gas classes: o H2S, S2, SO2, SO3, H2O, H2, CO2, CO, N2O, NO, NO2, CH4, C2H2, C2H4, C2H6, and other hydrocarbons, CH3OH and other potential alcohols o Measurement of all of these compounds must take place within the same sample, without switching detectors required (measurement must be simultaneous) • Software and manuals to utilize the equipment accessible on Windows (Windows 10 preferred), or with computers provided o If bidder is only providing software, it must be Windows 10 compatible, but if bidder is providing computer, they must provide the operating system required. • Electronics must be UL listed