Specifications include, but are not limited to: 2.2.1.Must meet all requirements of the EPA Method EPA 533/537/537.1 with a demonstrated ability to limit interference to less than 1/3rd the method reporting level for each of the analytes identified therein. Method performance must be documented and fully supported by the vendor 2.2.2.Must include a fully supported, established method for PFAS extraction 2.2.3.Must have the ability to extract 20 mL - 4 L aqueous samples 2.2.4.Must have the ability to simultaneously extract up to 6 samples with a single instrument 2.2.5.Must allow the use of up to 5 different solvents for the conditioning, rinsing and elution of samples 2.2.6.Must minimize Teflon® components, using PEEK material in the sample and solvent flow paths 2.2.7.Must accommodate cartridges with 1 mL, 3 mL or 6 mL plungers 2.2.8.Must be a new instrument (not refurbished) 2.2.9.Must separate organic and aqueous waste into separate containers using separate waste lines 2.2.10. Must not require operation in a fume hood