Specifications include, but are not limited to: 1. Autosampler with protective cover, with a minimum capacity of 50 samples. 2.Start Up Kits to Allow Operation upon Start up, including pump tubing and gas connections lines / tubing, and sample tubes. 3.Computer, Monitor & Software to run Instrument and process data. 4.Vacuum Pump 5.Chiller/Heat Exchanger. 1. Must be able to operate in temperatures ranging from 15 degrees C to 30 degrees C. 2. Must be able to operate in humidity ranging from 30-80%. 3. Must be able to exceed requirements for analysis outlined in 40 CFR 136 appendix B, 40 CFR 141.23, and EPA method 200.8, rev 5.4. 4. Exhaust flow must not be greater than 200 cfm. 5. Electrical requirements must be compatible with 120V outlets. 6. Software must be capable of uploading data into StarLIMS. 7. Installation Completed by Supplier. 8. Must Include 3-4 days of In-House Training of Instrument and Software Operation.