A. Refurbished / Reconditioned Mammography System System must be equivalent to the Hologic Selenia Dimensions 3D Mammography System with tomosynthesis, C-View, and CAD capability. Model Year Requirement: The mammography system must be a 2014 model year or newer. Vendor must provide documentation verifying the original manufacturing year. All electrical components must be fully refurbished and tested to OEM specifications. Cosmetic components may be reconditioned but must be free of cracks, dents, warping, or shielding defects. Vendor must provide the following refurbishment documentation at delivery: refurbishment checklist; electrical safety testing results; component replacement list; and final OEM-equivalent QC and calibration results. The system must be installed to maintain full functional compatibility with RCC’s electrical and imaging environment. B. Software and Functional Capabilities Must support digital breast tomosynthesis (3D), 2D FFDM imaging, C-View synthesized 2D imaging, diagnostic and screening workflows, and dynamic tube head motion. Must include all required legal, valid, and transferable software licenses. Software must support all DICOM and IHE integration requirements listed in Section E. C. X-ray Gantry Requirements Constant-potential, high-frequency generator equivalent to OEM specifications. Tungsten bi-angular X-ray tube with high heat capacity. Rhodium, silver, and aluminum filtration. Auto-retracting linear anti-scatter grid. Gantry mechanical movement must meet OEM integrity and safety specifications. D. Acquisition Workstation Multi-core Intel processor. Minimum 16 GB RAM. Minimum 2 TB internal storage. NVIDIA GPU capable of tomosynthesis reconstruction. Windows 10 64-bit operating system. Must include: 1.2 MP technologist control monitor; pull-out keyboard drawer; keyboard and mouse; technologist X-ray shield. Must include all OEM-required imaging applications, drivers, and connectivity software. E. Connectivity and PACS Integration Must fully support the following DICOM services: Modality Worklist (MWL), Storage, Storage Commitment, Query/Retrieve, Print, Modality Performed Procedure Step (MPPS), and Radiation Dose Structured Report (RDSR). Must support IHE Scheduled Workflow and Mammography Profiles. System must be fully compatible with IQ-View PACS. Vendor must demonstrate successful transmission of test images to RCC PACS during installation.