This requirement is for Olympus Vanta CORE XRF Metals Analyzer OR equal.
"Or equal" items shall meet or exceed the following qualities:
a. Ability to instantly identify exact chemical composition of metals containing Mg, Al, Si, P, S, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, W, Hf, Ta, Re, Pb, Bi, Zr, Nb, Mo, Pd, Ag, Cd, Sn, and Sb.
b. MIL-STD 810G drop test (4 Foot drop)
c. Self-Calibration on a known standard.
d. IP54 dust and water protection
e. Solid detection shutter to prevent possible damage.
f. Battery: Rechargeable Li-ion battery with hot-swap functionality to maintain power during battery change
g. Accessories: Waterproof Carrying Case, two (2) Li-ion Batteries, Docking Station, 316 Stainless Steel Calibration Check Reference Coin, Ten (10) Spare Windows, Integrated Wrist Strap, DELTA PC Software, Factory Authorized Training & Support
h. Environmental Temperature Range: -10 C to 50 C (14 F to 122 F)
i. Data Transfer: USB
NOTE: Once the solicitation is posted, it is incumbent upon the interested parties to review this site frequently for any updates/amendments to any and all documents. Any prospective contractor must be registered in the System for Award Management (SAM) (https://www.sam.gov/) at the time the quote is submitted in order to be eligible for award.