Contractor shall provide the University the following: 151 Class II Biological Safety Cabinets: With the following containment tests and certifications to be performed: HEPA Filter Leak Test Airflow Smoke Patterns Down flow Velocity Profile Test Inflow Velocity Test Site Installation Assessment 37 Laminar Flow Hoods (Clean benches, PCR hoods, Chem hoods with HEPA supply): Test and Certifications 1 Biological Safety Cabinet (Class II): ANSI Z 9.5 Test 1 “Walk-in” Biosafety Cabinet: Test and Certification 1 Animal Bedding Disposal Station: Test and Certification 11 Animal Cage Rack Systems: Test and Certification 6 Insect Lab Filter Banks: Test and Certification of Filter Banks 2 Animal Lab Filter Banks: Test and Certification of Filter Banks 2 BSL3 Building Filter Banks (bag in/bag out) – Test and Certification of Filter Banks