NMR (nuclear magnetic resonance) specifications for use in the Seafood Laboratory 1: The system must be equipped with the true Broadband capable of tuning between the full range of frequencies of 29Si to 31P and be installed pre-configured for 1H, 19F, 13C, 31P, 23Na, 11B, 7Li, 29Si, 27Al, 59Co. 2: The system must have the ability to manually tune and match to the sample to gain maximum SNR on the broadband channel, as well as have two nuclei on the broadband channel that can be duel tuned for auto switching. 3: The system must be equipped with 3 Axis pulsed field gradients 4: The system must be equipped with pulsed field gradients equal to or greater than 0.5 Tm-1 5: The magnet temperature must be stabilized within <0.0010C for optimum shim stability 6: The probe must be fully removable by the end user for cleaning and easy access for custom research modifications.