Mass and Spectral Accuracy (With optional Accurate Mass Package) Mass Accuracy – 1 µL injection of 100 pg/µL OFN, scanning from 50 to 300 u, will give its monoisotope at at m/z 271.987 ±0.005 Spectral Accuracy - 1 µL injection of 100 pg/µL OFN, scanning from 50 – 300 u, spectral accuracy will be 99.0% Nominal mass data may sometimes be inadequate to identify or confirm a compound, necessitating a measurement with increased mass and spectral accuracy.; Stainless Steel Ion Source Helium Carrier Gas Sensitivity, Scan Mode 1 pg OFN >= 250:1 S/N scanning from 50-300 u at nominal m/z 272 ion demonstrated at install Sensitivity is critical to meeting and/or exceeding analysis requirements especially for low detection level. For scan sensitivity S/N measurement the scan range must be reflective of normal operation parameters.; Ultra Inert Column of conventional size and Ultra Inert inlet liner The installation check out column should be the conventional dimension 30m x 0.25mm I.D. x 0.25µm for Helium Carrier Gas Ultra Inert Column, which the Manufacture offers the MS Library for major GC/MS applications using that column. The inlet liner should be Ultra Inert for demonstration of installation check out. The system can be “ready- to-go” after the installation for most GC/MS users without venting MS.