Specifications include, but are not limited to: 4.2.1 Detection Method – High temperature combustion at a minimum of 680°C utilizing infrared detection 4.2.2 Measuring Range – 0.05 to 500 mg/L TOC 4.2.3 Measuring Time – Capable of measuring two replicates for Total Carbon in 10 minutes or less 4.2.4 Power Supply – 110 VAC, 50/60 HZ 4.2.5 Autosampler Capacity – A minimum of 75 positions 4.2.6 Autosampler Sparging – Sparging with carrier gas of the next vial to be sampled 4.2.7 Autosampler Mixing – Effective mixing of the current vial being sampled 4.2.8 Operational Mode – PC control of analyzer utilizing a high-end processor and memory configuration; Monitor with a resolution of 1920 x 1080 or more 4.2.9 Operating Software – Windows 10; Controls all instrument functions; capable of importing sample ID’s from text files into tray tables and exporting to LIMS/Excel 4.2.10 Documentation – Vendor shall provide PM schedule and PM instructions 4.2.11 Documentation – Vendor shall provide a list of consumable products and critical spare parts list.