Specifications include, but are not limited to: 1. Unit Requirements • 2 Point Technique (kV – mAs) • kV Range: 40 – 130kV (1kV settings) • mA Range: 70 – 400mA (based on KV steps) • Time Range (based on mAs selection): 0.001 -3 seconds • Exposure hand switch • 300kHU X-Ray Tube, Rotating Anode • Focal Spot: 0.6mm (Small) / 1.2mm (Large) • Manual collimator: High brightness LED, 30 second timer (adjustable), Collimator rotation 120° • Motorized movement • Battery powered (with rechargeable batteries) • Weight: Up to 1000 lbs • Max height: 90" • Flat Panel Detector and grid housing 2. Arm Extension (Telescopic) • 14" to 16" • Front range: 28" to 44" • Lateral range: 20" to 36 3. Rotation of X-Ray Tube Head • A-swivel, ±180° • B-swivel, ±320° (from transport position) • C-swivel, 127° (90° forward, 37° backward) 4. Workstation Hardware & Software • Onboard PC with minimum specs: 8GB RAM, 500GB SSD, Windows 10 Professional 64-bit, LAN & WLAN connectivity • Touchscreen Display • Intuitive Technologist Exposure Guidance • Advanced Image Processing and Algorithms • Image Reject Tracking and Analysis • CD/DVD/USB Drive Export/Output • DICOM Store, Print, and Modality Worklist • Three (3) Year Warranty