Specifications include, but are not limited to: Required; Electrical; • 120 Volt, 50/60 Hz (Controls) • 480 Volt, 50/60 Hz, 3-Ph (Steam Generator/Vacuum Pump); Autoclave/Steam Sterilizer • Recessed, to fit 28 ¼" (W) x 74 ¼" (H) x 45 ½" (L), vendor to measure/verify during site visit • Seismic tie-down • Prevacuum model • Gauges, visible, for jacket and chamber pressure • Capable of measuring/monitoring temperature in the sterilizing chamber • Safety features o Control lockout o Chamber float switch/alarm o Pressure relief valve o Emergency stop • Chamber o Volume/capacity 250L or similar o Stainless steel "316L” o Meets ASME section VIII standard o Insulated, water- and oil-resistant • Steam generator o Integral, electric, carbon steel o Auto flush • Door, vertical sliding, power door • Loading cart with transfer carriage and chamber, track assembly • Control system o Touch screen o Cycle values and operating features adjustable and programmable (including sterilization temperature and time) o Cycle values verifiable prior to cycle operation o Automatic/programmable startup/shutdown of autoclave at beginning/end of day o Cycle data saved internally, can be retrieved/viewed/exported o Mobile/PC alerts