• Each cylinder must have specific labels attached (See Page 26). • Compatible with the DataMaster DMT-G breath test instrument manufactured by "National Patent Analytical Systems, Inc./ and the intoximeter DMT-G breath test instrument manufactured by lntoximeters. • 105-liter non-refillable cylinder. • 208 parts per million (ppm) Ethanol in Nitrogen gas mixture at sea level (pressure of 760 mm of HG) when used with breath alcohol devices (an ethanol content equivalent to 0.08 g/210L ethanol breath standard (+/- 2%). • National Institute of Standards and Technology (NIST) traceable. • Certified content and ethanol vapor concentration with a certificate of analysis provided for each lot received. • Minimum two-year reliability from date of manufacture. • Expiration date and lot number printed on each cylinder. • Manufacturer is accredited to the following standards with scopes that include the manufacturing, testing & calibration of NIST traceable dry gas ethanol reference materials: ISO/IEC 17025 & ISO/IEC 17034. • Lots shipped to Alaska must have been manufactured no more than sixty days prior to shipment. • Each shipment will contain only one lot per batch of hundred cylinders. • Pressure of each tank is 1,000 PSI or greater (at room temperature) when received. • Each tank of a lot is tested & verified by manufacturer before each shipment. • Sample labels are to be approved by Crime Lab.