Specifications include, but are not limited to: 1. Measuring sensor must be linear variable differential transducer (LVDT) 2. Resolution of at least 1.25nm/digit 3. System must allow expansion measurement between ambient and 1600°C with optional low temperature attachment to -150°C 4. System shall be vacuum-tight by design (10¯5 mbar) for tests on materials sensitive to oxygen 5. The measuring compartment must be made of aluminum or fused silica 6. The controlled heating and cooling rates must be adjustable between 0 and 400°C/s 7. User-exchangeable Dilatometer measuring systems should be available 8. Data acquisition with up to 1000 samples 9. Dilatometer Measurement and Evaluation Software for CCT, CHT, and TTT diagram generation must be supplied with the system 10. Tensile and compression force up to 25kN 11. Deformation measuring range 100µm up to 5000µm 12. Must have DSC measuring system capable to 1450°C including a susceptor