Specifications include, but are not limited to: X-ray system and gantry that shall meet all the relevant equipment minimum requirements specified in the Mammography Quality Standards Act FDA-approved 2-D and 3-D (tomosynthesis). FDA-approved for 2-D synthesized images from 3-D tomosynthesis data (Generates 2-D image from 3-D tomosynthesis data). System shall filter back super resolution reconstruction for synthesized images Gantry height and angle movement shall be motorized and movement locks shall be included Unit shall have a means to automatically select the optimum exposure factors, depending on the x-ray absorption characteristics of the breast. Automatic exposure control for all breast types including implants. Unit should include accessories required to produce images of at least 1.8X magnification.