Specifications include, but are not limited to: FEATURES - Unit shall meet or exceed the following: 2.1.1 Unit will comply with AAMI (Association for the Advancement of Medical Instrumentation) Standards. 2.1.2 Unit will be easy to use with instructions clearly posted on unit. 2.1.3 Unit will measure blood pressure and pulse rate. Units with weight measurement systems will not be considered. 2.1.4 Unit will have a fully automated adult size upper arm pressure cuff. 2.1.5 Unit will have LED display of readings. 2.1.6 Unit will have external meter for tracking usage. 2.1.7 Unit will have posted blood pressure guidelines as defined by current medical industry standards established by organizations such as the Centers for Disease Control (CDC). 2.1.8 Unit will at a minimum, possess four (4) information display panels for dissemination of health and wellness information. 2.1.9 Unit will utilize oscillometric methodology. 2.1.10 Unit will inflate at approximately 8mmHg/Sec. 2.1.11 Unit will operate accurately under temperature conditions ranging from 50 to 104 degrees F (10 to 40 degrees C). 2.1.12 Unit will operate accurately under atmospheric humidity conditions ranging from 10 to 95%. 2.1.13 Unit must be capable of readings as follows: A. Systolic: +50 to +260 mmHg B. Diastolic: +30 to +240 mmHg C. Pulse: 40bpm to 199 bpm 2.1.14 Test time will be approximately 1 minute or less. 2.1.15 Unit will possess an emergency stop mechanism with rapid deflation when pressure reaches overpressure limit of 310 +/- 10 mmHG. 2.1.16 Physical Dimensions shall be approximately: A. Height 47.8 to 68.0” B. Width: 24.2” C. Depth: 20.3” D. Weight: 82 to 90 pounds. 2.1.17 Input power requirements at a minimum: 117 VAC, 60HZ, 2.5 amps with a 3-wire grounded outlet. 2.1.18 Unit shall be adaptable for mobility impaired individuals. 3.1 The blood pressure stations will be installed and maintained by the Contractor. 3.2 Locations will be designated by City and are noted below as Section 4.0. 3.3 Locations and units may be added or deleted as necessary. 3.4 The blood pressure stations no longer needed by the City or at contract termination will be removed within 30 calendar days of notification to Contractor