Specifications include, but are not limited to: Provide (2) two digital radiographic imaging units Dicom compliant 65 KW Generator Perform digital imaging at separate table and wall bucky detector (2 detector unit) one fixed and one wireless, detection area of minimum 17”X17” Capable of performing upright scannogram images of spine or pelvis to ankle Motorized elevating vertical table with float top capabilities (2) sets of maintenance and repair manual(s) (2) sets operators Manual(s) Remote diagnostic capability X-Ray tube on floor mounted rail configuration Annode Heat Storage Capacity Wall Detector/Multipurpose Automated or manual driven positioning Full interface to existing hospital PACS and RIS systems Minimum hard disk, 500 Gb Minimum RAM storage capacity, 4 Gb DVD/CD writer Monitors size minimum 19” Mains voltage 115V 230 V Fuse 10A Floor mounted height adjustable table Up to 220 lbs patient weight