Specifications include, but are not limited to: Line 1: Capillary Electrophoresis Instrument Referenced Brand/Product No.: Life Technologies Corp. AB 3500XL, No. 4406016 Quantity: 1 each Delivery: Austin, TX 78752 A. 24 Capillary with Array and Polymer B. Capillary arrays: The internally uncoated capillaries are supplied in assemblies of 24 capillariesper array, with a built-in frame. C. DNA sequencing and/or fragment analysis reagents and consumables for system qualification. D. Laser: 1. Single line 2. 505 nm solid-state laser excitation source E. Electrophoresis Voltage: 20 kV Limit F. Oven Temperature: Active control from 18°C to 70°C G. Polymer Pump H. Performance Optimized Polymer (POP) Pouch 1. POP-4 2. POP-6 3. POP-7 I. Anode Buffer Container (ABC) J. 96- and 384- Well Plates K. Cathode Buffer Container (CBC) L. Operating Environment 1. Temperature 15°C-30°C (room temperature should not fluctuate more than +/-2°C) M. Consumables for system qualification N. Current Limit: 15 amps O. Maximum Power Dissipation: 417 VA, 371 W (not including computer and monitor) P. Main power voltage 1. 100-240V +/- 10% 2. 50-60 Hz +/- 10% Q. Dimensions 1. Width (Closed Door): 61 cm 2. Width (Open Door): 122 cm 3. Depth: 61 cm 4. Height: 72 cm R. Weight: 82 kg