Specifications include, but are not limited to: 1. EL-USB-TP-LCD+ Data Logger with LCD Screen. Lascar Electronics EL-USB-TP-LCD+ or equal. 32K reading memory with LCD external temperature range -40C to +125C +/- 1 deg. accuracy. Power supply 3.6 1/2AA lithium battery with collar seal. Meets CE, RoHS Government Regulations. Logger must have custom delay firmware for pharmaceutical refrigerator and freezer use. Unit must be compatible with Easy Log software. The logger itself must be able to be affixed to the external wall of the refrigerator and or freezer by magnetic or similar means. A sensor probe in glycol must be provided with the data logger and have a National Institute of Standards and Technology (NIST) two (2) year calibration certification from an ILAC/MRA certified laboratory. 2. EL-WiFi-TP+ WiFi Temperature Probe Data Logger. Lascar Electronics EL-WiFi-TP+ or equal. A temperature probe will be supplied in glycol with the sensor. It measures the temperature in which the probe is situated. Data is transmitted wirelessly via a WiFi network to a personal computer and cloud environment and viewed using an Easy Log software. 32K reading memory with LCD external temperature range -40C to +125C +/- 1 deg. accuracy. Power supply 3.6 1/2AA lithium battery with collar seal. Meets CE, RoHS Government Regulations. Unit must be compatible with Easy Log software. The logger sensor itself must be able to be affixed to the external wall of the refrigerator and or freezer by magnetic or similar means. A sensor probe in glycol must be provided with the sensor and have a National Institute of Standards and Technology (NIST) two (2) year calibration certification from an ILAC/MRA certified laboratory. 3. TEKCELL/TC SB-AA02 Lithium Battery. TEKCELL/TC SB-AA02 or equal. Replacement battery for EL-USB-TP-LCD+ Data Logger, 3.6 1/2AA Lithium battery with custom collar seal on positive end. 4. Data Logger Calibration. If required and purchased by ODH, the Contractor shall provide two (2) year calibration for new Data Loggers, purchased under this contract, and existing, State-owned data loggers. Must include NIST certification from an International Laboratory Accreditation Cooperation/Mutual Recognition Arrangements (ILAC/MRA) certified laboratory. Re-certification labels must be placed on each device after completion. Electronic certificate must be available for each device. The units will be shipped for calibration by the Ohio Department of Health to the Contractor at the Ohio Department of Health’s expense and returned by the Contractor at the Contractor’s expense to: Ohio Department of Health, 900 Freeway Drive North, Building 8, Columbus, OH 43229. 2-Year certification is required. Must meet NIST Standards. Calibration and return will be within 30 days of receipt by the Contractor. 5. Glycol Probe with Wire. High accuracy, 3 meter long thermistor type A probe in sealed Glycol filled bottle for vaccine monitoring; measurement range: -40°C to +60°C / -40°F to + 140°F; Accuracy - ±0.1°C / ±0.2°F. 6. Replacement Parts. Any replacement parts needed for new Data Loggers, purchased under this contract, and existing, State-owned data loggers. 1. Styrene Shipping Containers with Lids for vaccines. Polar Tech or equal. 1 ½” wall thickness, with insert lid, UPS Shipper approved. The storage capacity of interior space cannot be less than that specified. However, sizes offered may be larger by no more than one (1) inch in any direction. The following sizes are required: 8” LG x 6” W x 4” D 204F Polar Tech or equal 13” L x 13” W x 12” D 240F Polar Tech or equal 19” L x 12” W x 16” D 266F Polar Tech or equal 8” L x 6” W x 9” D 207F Polar Tech or equal 17” L x 10” W x 12” D 247F Polar Tech or equal 22” L x 16” W x 16” D ON69F Polar Tech or equal 12” L x 10” W x 7” D 227F Polar Tech or equal 19” L x 12” W x 12” D 261F Polar Tech or equal 25” L x 16” W x 12” D 295F Polar Tech or equal