Specifications include, but are not limited to: UPS Powervar with capacity for LCMS-8045 plus nitrogen generator. The unit includes a 5.2KVA on-line conditioned UPS at 208/230/240 Volts input @ 60 Hz and 120/208/230/240 Volts output @ 25 amps and 60Hz. System includes low impedance isolated transformer with surge suppression and filtering network. The power conditioner is RoHS compliant. Case size is 28.9"H x 11.8"W x 31.91"D. Shipping weight is 395 lbs. Input plug (1)L6-30P and output (1)L6-20R, (4)5-20R Receptacles. UPS Powervar with capacity for LCMS-8045 only. The unit includes a 4KVA power conditioner with 208/230/240 Volts input @ 60 Hz and 120/208/230/240 Volts output @ 20 amps and 60Hz. System includes low impedance isolated transformer with surge suppression and filtering network. The power conditioner is RoHS compliant. Case size is 28.9"H x 11.8"W x 31.91"D. Shipping weight is 299 lbs. Input plug (1)L6-30P and output (1)L6-20R, (4)5-20R Receptacles.