Specifications include, but are not limited to: General Operation Available mAs at 120 kV Indicate detector and cell size Indicate scan coverage Gantry weight limit and maximum length Gantry maximum tilt angle Is the gantry speed variable Indicate maximum scan length Indicate maximum reference scan length The system must have radiation dose optimization/reduction capabilities (describe) Include a dose profile for a scan using 120kV, 200 mA & 1 sec scan time for each available slice thickness Clinical Applications Indicate frames per second while performing CT fluoro Image Quality Indicate the maximum image resolution Indicate display rate while performing a 20 sec helical acquisition Indicate image reconstruction time based on a 0.5 sec scan Indicate low contrast resolution in % at 0.3% using 120kV, 400 mA & 1 sec scan time and a 5 mm slice thickness Indicate spatial resolution in lp/cm at 2% MTF using 120kV, 400 mA & 1 sec scan time and a 2 mm slice thickness Indicate noise standard deviation in mm using 120kV, 400 mA & 1 sec scan time and a 5 mm slice thickness Tube Specifications X-ray tube has a history of low tube failure Tube has the ability to handle rapid volume acquisition or in between successive acquisitions in a busy, high patient throughput facility Indicate heat storage capacity and cooling rate of x-ray tube